E-Series Horizontal Laminar Flow Hoods

OVERVIEW  |  E-SERIES VERTICAL FLOW  |  E-SERIES HORIZONTAL FLOW
TABLE-TOP LAMINAR FLOW  |  C-FRAME WORKSTATION  |  CUSTOM SERIES
TABLES  |  OPTIONS FOR CLEAN BENCHES & WORKSTATIONS

horizontal laminar flow hoodsGeneral Information
A self contained SAM Fan Filter Unit provides Class 100 HEPA filtered air in a horizontal laminar flow pattern within the workspace.

Electrical
Standard voltage is 115 volts with other voltage options available. Separate On/Off switches for the fan and lights are located above the workspace. All 115 volt models are equipped with a 8 foot power cord and 3 – prong plug. Other voltages require electrical connection to the power source by a qualified electrician.

Final Filter
HEPA filters are 99.99% efficient @ 0.3 micron or larger with anodized  aluminum frame. ULPA grade filters are 99.9995% efficient @ 0.12 micron, and are available as an option.

Prefilter
A pleated type, disposable prefilter rated at 30% ASHRAE efficiency is standard.

Speed Control
A solid state variable speed control is mounted on the SAM unit.

Grille
Horizontal flow clean benches have a polystyrene grille inserted in an anodized aluminum frame which is hinged to open. This permits removal of the HEPA filter without removing the SAM unit from the frame.

Construction
Clean Bench frames are white painted steel tubing with clear acrylic side panels and adjustable leg levelers. Work surfaces are solid core with white high pressure laminate finish.  ESD, perforated and stainless steel work surfaces are available as options. Clean Bench models are available in a sit-down style with work surface 30” from floor or stand-up style with work surface 40” from floor.

Lighting
All horizontal flow models with lights are equipped with a 4-lamp light fixture, 120 volt, 60 hz with T8, 32W electronic ballast. Fluorescent light tubes are not furnished with the units because of potential breakage during shipment.

Shipping
Horizontal Flow Clean Benches are fully assembled.

Options
See a following page for options and ordering information.

Ordering Information
Sit-down Style Horizontal Flow w/Lights and Work Surface 30” from Floor
Model No. Voltage Nominal Work Surface Overall Dim. Opening Size Weight
13255 115V 2 Ft. 24″ x 25″ 26”L x 44”D x 66”H 24”L x 30”H 275 lbs.
13243 115V 4 Ft. 48″ x 25″ 50”L x 44”D x 66”H 48”L x 30”H 360 lbs.
13244 115V 4 Ft. 48″ x 32″ 50”L x 51”D x 66”H 48”L x 30”H 375 lbs.
13245 115V 6 Ft. 72″ x 25″ 74”L x 44”D x 66”H 72”L x 30”H 490 lbs.
13246 115V 6 Ft. 72″ x 32″ 74”L x 51”D x 66”H 72”L x 30”H 505 lbs.
13247 115V 8 Ft. 96″ x 25″ 98”L x 44”D x 66”H 96”L x 30”H 630 lbs.
13248 115V 8 Ft. 96″ x 32″ 98”L x 51”D x 66”H 96”L x 30”H 650 lbs.

NOTE: Fluorescent light tubes not included

Ordering Information
Stand-up Style Horizontal Flow w/Lights and Work Surface 40” from Floor
Model No. Voltage Nominal Work Surface Overall Dim. Opening Size Weight
13256 115V 2 Ft. 24″ x 25″ 26”L x 44”D x 76”H 24”L x 30”H 240 lbs.
13249 115V 4 Ft. 48″ x 25″ 50”L x 44”D x 76”H 48”L x 30”H 390 lbs.
13250 115V 4 Ft. 48″ x 32″ 50”L x 51”D x 76”H 48”L x 30”H 410 lbs.
13251 115V 6 Ft. 72″ x 25″ 74”L x 44”D x 76”H 72”L x 30”H 515 lbs.
13252 115V 6 Ft. 72″ x 32″ 74”L x 51”D x 76”H 72”L x 30”H 530 lbs.
13253 115V 8 Ft. 96″ x 25″ 98”L x 44”D x 76”H 96”L x 30”H 700 lbs.
13254 115V 8 Ft. 96″ x 32″ 98”L x 51”D x 76”H 96”L x 30”H 730 lbs.

NOTE: Fluorescent light tubes not included

Options
208-230V, 50/60 hz motor Casters with brakes ESD Acrylic side & back panels
277V, 60 hz motor Duplex electrical outlets ESD work surface
250V, 50 hz lighting for export models Front Face Shields Stainless steel work surface
Duct Collar Adaptor, 10” or 12” diameter Magnehelic Gage Static Control Package
ULPA Filter 99.9995% eff. @ 0.12 micron or larger Perforated work surface

Refer to following pages for more options